DA 2000(08) - Breast Implants: NovaGold

Document details:

Type: Medical Device Alert
Series No: DA 2000(08)
Audience: Healthcare professionals
Published:
Format:
Size: A4
Pages: 4
Price: free
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Please note: In October and November 2010, the MHRA moved from its Market Towers address to a new location on Buckingham Palace Road. Please go to the contact us section of the website for updated contact details.

Please note that contact names and/or numbers may have changed since this notice was published. If you need to contact us about a particular notice,  email dts@mhra.gsi.gov.uk quoting the reference number and title of the notice.
_________________________________

December 2000

MANUFACTURER/SUPPLIER
Novamedical/Somatech Medical Ltd

PROBLEM
Recall of NovaGold™ breast implants

For immediate attention by:

Health Authorities (England) - Chief Executives
NHS Trusts (England) - Chief Executives
Primary Care Trusts (England ) - Chief Executives

IMMEDIATE ACTION

1. Do not implant NovaGold™ breast implants.

2. Identify and isolate all stocks of NovaGold™ breast implants and return them to the supplier.

3. Report any adverse incidents concerning implanted devices to MDA.

DISTRIBUTION REQUIRED
Please bring this notice to the attention of all who need to know or be aware of it. This will include distribution by:

TRUSTS to:

  • Liaison Officers (for onward distribution)
  • Medical Directors
  • Plastic Surgeons and all Surgeons involved in breast reconstruction
  • Directors of surgical units involved in breast reconstruction
  • Nurse Executive Directors
  • Specialist Nurses involved in breast cancer care

HEALTH AUTHORITIES to:

  • Liaison Officers (for onward distribution)
  • Registration Inspection Units
  • Chairs of Primary Care Groups
  • General Medical Practitioners - for Information in case women contact you
  • Practice Nurses - for Information in case women contact you
  • Private hospitals and clinics involved in plastic, cosmetic and breast reconstructive surgery

PRIMARY CARE TRUSTS to:

  • Chief Executives (for onward distribution)

BACKGROUND
NovaGold™ breast implants are made in Germany and are supplied by Somatech Medical Ltd in the UK. Approximately 250 women in the UK have been implanted with these products since they were first introduced in 1996.

NovaGold™ breast implants have a silicone elastomer shell and are filled with a polyvinylpyrrolidone (PVP) hydrogel and guar gum gel. PVP was previously used in the filler of another breast implant (MistiGold) manufactured by Bioplasty Inc which were supplied until the mid 90's.

Hydrogels are polymeric materials that have the ability to swell in water without dissolving and to retain water within their structures. There are many different types of hydrogels which are used in a wide variety of products including, pharmaceuticals, foods and medical devices such as contact lenses and surgical dressings. The information given in this Device Alert applies only to NovaGold™ breast implants and not to other uses of hydrogels.

CONCERNS
Because of concerns raised in relation to breast implants, MDA has been undertaking a series of planned investigations into the safety of breast implant fill materials. Only one adverse incident (an implant rupture) has been reported involving NovaGold™ breast implants. However, MDA's review has identified that the manufacturer's biological safety assessment of this product in inadequate, due to the lack of long-term toxicity data or clinical follow-up, together with methodological flaws in some of the pre-clinical tests. The particular concerns are:

  • uncertainty over the metabolic fate of the filling material;
  • pathological changes suggesting a systemic effect in rabbits following implantation of the filling material for 28 days. No long-term toxicity data are available to allow an assessment of the significance or reversibility of these effects.
  • lack of data to assess risks from the implantation of guar gum.

In light of this, MDA has concluded that, as a precautionary measure, no more NovaGold™ breast implants should be implanted until these concerns have been addressed.

ADVICE FOR IMPLANTED WOMEN
Women who are worried should be offered a consultation. It should be emphasised that a definite risk has not been identified. The concern lies only with the adequacy of the safety assessment for NovaGold™ breast implants. There is no evidence at present to recommend explantation of NovaGold™ breast implants. Any decision to explant NovaGold™ breast implants should be taken jointly be the woman and her clinician/surgeon, based upon the individual clinical circumstances, balancing the potential risk arising from the lack of safety data against the known risks associated with any surgical procedure.

At present, there is no information to indicate that there are any specific risks to women or any effects on children, either prior to or during pregnancy or from breast feeding.

There is no indication from the scientific studies currently available to suggest that pregnancy should be postponed. Women who are concerned about the inadequacy of the safety assessment may wish to consider waiting until further safety tests have been performed and evaluated, before planning a pregnancy.

Annex 1 has been written to provide information to women. In addition they can ring the Health Information Service on 0800 665544 or, where it is available NHS Direct on 0845 46 47. (In Scotland, the Scottish NHS Helpline on 0800 224488).

PAYMENT FOR CONSULTATION
The Department of Health is currently discussing reimbursement for clinical consultations with the manufacturer.

FUTURE INVESTIGATION/ADVICE
The manufacturer has agreed to carry out further testing and to provide the results to MDA for review. MDA's analysis of the test results will be made available along with clear recommendations if appropriate.

ENQUIRIES
Enquiries to the company should be addressed to:

Dr Aydin Dogan
NOVAMEDICAL Products GmbH
Heinestr. 3
D-40789 Monheim
Germany

Tel: +49 21 73 95 38 0
Fax: +49 21 73 95 38 55

Enquiries to the MHRA should quote reference number 20001109.015-1.


© Crown Copyright 2000

----------------------------------------------------------------------------

ANNEX 1

INFORMATION FOR WOMEN ABOUT NovaGoldTM BREAST IMPLANTS

The Medical Devices Agency (MDA), which is part of the Department of Health, has provided this information sheet. It is designed to give you information about NovaGold™ breast implants.

Background
NovaGold™ breast implants are filled with a hydrogel. Hydrogels consist mainly of water with a substance to give them a "jelly-like' consistency.

There are many different types of hydrogels used in a wide range of products including foods, medicines and contact lenses. MDA's advice relates only to the hydrogels used in the NovaGold™ breast implants on sale in the UK, and not to hydrogels used in other products.

Reason for removing NovaGold™ breast implants from the UK market

It should be emphasised that a definite risk has not been identified. There are no known cases of harm caused by NovaGold™ breast implants. The concern is only with the way the safety of the hydrogel filler has been assessed.

MDA has looked at the breast implants filled with hydrogel on the UK market. The information provided by the makers of NovaGold™ breast implants was looked at and MDA found that there is not enough information to fully assess the filler material.

Because of their concerns, MDA has decided that no more NovaGold™ breast implants should be used until more information is available.

MDA is not recommending the removal of NovaGold™ breast implants.

If new information comes to light which changes MDA's advice they will immediately inform doctors and women about this.

General Advice
If you think you have NovaGold™ breast implants and you are worried you may wish to:

  • contact your surgeon or your doctor,
  • get advice from the Health Information Service on tel. 0800 665544 or, where available, NHS Direct on tel. 0845 4647. (in Scotland, the Scottish NHS Helpline on 0800 224488)

Removal - At present there is no information to suggest that you should have your NovaGold™ breast implants removed. If you have concerns discuss these with your surgeon or doctor.

Rupture - If your NovaGoldTM breast implants have ruptured, you should discuss this with your surgeon.

Having a baby or breast feeding - At present there is no information to indicate that there is any actual risk of effects on children, either prior to or during pregnancy or from breast feeding.