| Type: | Medical Device Alert |
|---|---|
| Series No: | MDA/2012/079 |
| Audience: | Healthcare professionals |
| Published: | 20 November 2012 at 14:00 |
| Format: | Electronic only |
| Size: | A4 |
| Pages: | 4 |
| Price: | Free |
| ISBN/ISSN: | |
| Author: | |
| Copyright: | Crown |
CAS deadlines
Action underway: 11 December 2012, action complete: 7 January 2013
Note: These deadlines are for defibrillators to be identified and HeartSine contacted.
Ref: MDA/2012/079 | Issued: 20 November 2012 at 14:00
If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2012/079 or 2012/009/011/081/001
Technical aspects
Enitan Taiwo and Catriona Blake
Medicines & Healthcare products Regulatory Agency
Floor 4
151 Buckingham Palace Road
London SW1W 9SZ
Tel: 020 3080 7122/7219
Fax: 020 8754 3965
Email: enitan.taiwo@mhra.gsi.gov.uk
catriona.blake@mhra.gsi.gov.uk
Clinical aspects
Mark Grumbridge
Medicines & Healthcare products Regulatory Agency
Floor 4
151 Buckingham Palace Road
London SW1W 9SZ
Tel: 020 3080 7128
Fax: 020 8754 3965
Email: mark.grumbridge@mhra.gsi.gov.uk
Please report via our website: Reporting adverse incidents involving medical devices
Further information about CAS can be found on the CAS website (external link)
Alerts in Northern Ireland will continue to be distributed via the NI SABS system.
Enquiries and adverse incident reports in Northern Ireland should be addressed to:
Northern Ireland Adverse Incident Centre
Health Estates Investment Group
Room 17
Annex 6
Castle Buildings
Stormont Estate
Dundonald BT4 3SQ
Tel: 02890 523 704
Fax: 02890 523 900
Email: NIAIC@dhsspsni.gov.uk
Please report directly to NIAIC, further information can be found on the NIAIC website (external link)
Further information about SABS can be found on the SABS website (external link)
Enquiries and adverse incident reports in Scotland should be addressed to:
Incident Reporting and Investigation Centre
Health Facilities Scotland
NHS National Services Scotland
Gyle Square
1 South Gyle Crescent
Edinburgh EH12 9EB
Tel: 0131 275 7575
Fax: 0131 314 0722
Email: nss.iric@nhs.net
Health Facilities Scotland Incident Reporting and Investigation Centre (IRIC) (external link)
Enquiries in Wales should be addressed to:
Improving Patient Safety Team
Medical Directorate
Welsh Government
Cathays Park
Cardiff CF10 3NQ
Tel: 029 2082 3922
Email: Haz-Aic@wales.gsi.gov.uk
Samaritan public access defibrillator.
Model: PAD 300/PAD 300P.
Manufactured by HeartSine Technologies Ltd.
Specific serial numbers are affected.
Samaritan PAD 300/300P distributed from August 2004 to December 2010 with a warranted life of up to seven years.
The following serial number ranges are affected with one or both of the identified faults.
0400000501 to 0700032917 inclusive
08A00035000 to 10A00070753 inclusive
10C00200000 to 10C00210318 inclusive
Note: this range supersedes the one in the manufacturer’s FSN.
Risk of failure to deliver therapy caused by two different faults:
The manufacturer issued a Field Safety Notice
(471Kb) (FSN) in September 2012, providing advice on management of the defibrillator.
1. reserve Pad-Pak (battery) and instruction ‘hang tag’
2. the software upgrade, USB cable and installation guide.
Once the upgrade pack is received follow the instructions to ensure that:
Community defibrillation officers and staff responsible for the use, maintenance and purchase of these devices.
This MDA has been sent to:
Primary care trusts
CAS liaison officers for onward distribution to all relevant staff including:
Establishments registered with the Care Quality Commission (CQC) (England only)
This alert should be read by:
Please note: CQC and OFSTED do not distribute these alerts. Independent healthcare providers and social care providers can sign up to receive MDAs directly from the Department of Health’s Central Alerting System (CAS) by sending an email to: safetyalerts@dh.gsi.gov.uk and requesting this facility.
HeartSine Technologies Ltd
Canberra House
203 Airport Road West
Belfast
BT3 9ED
Tel: 028 9093 9400
Fax: 028 9093 9401
Email:data@heartsine.co.uk
Website: www.heartsine.com
If you have any comments or feedback on this Medical Device Alert please email us at: dts@mhra.gsi.gov.uk