MDA/2007/100 - Volumetric infusion pump. Smiths Medical (Graseby) 500 and 505 volumetric infusion pumps - all models

Document details:

Type: Medical Device Alert
Series No: MDA/2007/100
Audience: Healthcare professionals
Published:
Format: Electronic and paper
Size: A4
Pages: 4
Price: free
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Page 1

Issued: 20 December 2007 at 14:30
Ref: MDA/2007/100

 

 
Immediate action
Tick mark
Action
 
Update
 
Information request


 

Device:
Volumetric infusion pump. Smiths Medical (Graseby) 500 and 505 volumetric infusion pumps – all models.
 
Problem:
There is the possibility of delivering an unintended 1.7 ml bolus if the pump door is opened and then immediately closed following a stoppage of the pump.

>Page 2

Action by:
Users of these pumps.
 
Action:
To prevent unwanted boluses being delivered to a patient, always wait for a count of at least six seconds after opening the pump door before closing it again.
 

Distributed to:

NHS trusts in England - Chief Executives*
Commission for Social Care Inspection (CSCI) - Headquarters
Healthcare Commission (CHAI) - Headquarters
Primary care trusts in England - Chief Executives*

* via CE Bulletin

>Page 2

Contacts:
Details of manufacturer contacts and MHRA contacts for technical and clinical aspects.
Change of address or removal from address list for CSCI and Healthcare Commission.

>Pages 2-3

Appendix:
Smiths Medical Customer Information Bulletin

>Page 4



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 21 January 2008
Deadline (action complete): 20 February 2008

Page last modified: 20 December 2007