MDA/2007/098 - Enteral feeding pump. Fresenius Kabi Applix Smart pump - all models

Document details:

Type: Medical Device Alert
Series No: MDA/2007/098
Audience: Healthcare professionals
Published:
Format: Electronic and paper
Size: A4
Pages: 4
Price: free
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Page 1

Issued: 17 December 2007 at 11:00
Ref: MDA/2007/098

 

 
Immediate action
Tick mark
Action
 
Update
 
Information request

Device:
Enteral feeding pump.
Fresenius Kabi Applix Smart pump – all models.
 
Problem:
Misloaded or incorrectly fitted giving sets can permit free flow of feed to the patient with the possibility of fatal over-feeding.

>Page 2

Action by:
All medical, nursing and allied health professional staff involved with the management and use of these and other enteral feeding pumps.
 
Action:
Ensure that the giving set is correctly fitted in the lower tube guide and is slightly stretched in a straight line. See appendix for manufacturer’s instructions for use (in pdf only).

>Page 2

Distributed to:

NHS trusts in England - Chief Executives*
OFSTED  - Directors of Children’s Services
Commission for Social Care Inspection (CSCI) - Headquarters
Healthcare Commission (CHAI) Headquarters
Primary care trusts in England - Chief Executives*
Social services in England - Directors*

* via CE Bulletin

>Page 2

Contacts:
Details of manufacturer and MHRA contacts for technical and clinical aspects.
Change of address or removal from address list for CSCI and Healthcare Commission.

>Page 3

Appendix:
Extract from manufacturer’s instructions for use (in pdf only).

>Page 4



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 07 January 2008
Deadline (action complete): 11 February 2008