| Device: Enteral feeding pump. Fresenius Kabi Applix Smart pump – all models. |
|||||||||||||
| Problem: Misloaded or incorrectly fitted giving sets can permit free flow of feed to the patient with the possibility of fatal over-feeding. |
>Page 2 |
||||||||||||
| Action by: All medical, nursing and allied health professional staff involved with the management and use of these and other enteral feeding pumps. |
|||||||||||||
| Action: Ensure that the giving set is correctly fitted in the lower tube guide and is slightly stretched in a straight line. See appendix for manufacturer’s instructions for use (in pdf only). |