| Issued: 17 September 2007 at 14:30 |
Ref: MDA/2007/073
|
|
Immediate action |
|
|
Action |
|
|
Update |
|
|
Information request |
| Device: Electrosurgery bipolar unit: model TDB 60 manufactured by Eschmann Ltd (serial numbers with the last six digits from 5D 1201 to 7D-1273 inclusive are affected - see appendix for full list of serial numbers). |
|
||||||
| Problem: Failure of a component in the main control board may lead to the power output being up to four times that displayed on the front panel. As a precautionary measure, Eschmann are changing the main control board of affected units. |