| Issued: 15 August 2007 at 15:00 |
Ref: MDA/2007/064
|
|
Immediate action |
|
|
Action |
|
|
Update |
|
|
Information request |
| Device: Fluid warming set: Level 1 normothermic IV fluid administration and irrigation sets - D/DI and IR/IRI series manufactured by Smiths Medical. |
>Page 2 |
||||
| Problem: Small holes have been found in the inner lumen (aluminium tube) of some sets which can allow leakage of the heating fluid into the infusate with the potential for contamination. |
>Page 2 |
||||
| Action by: All those involved in the use of these devices including perfusionists, operating department practitioners, theatre staff/managers and anaesthetists. |
|||||
|
Action:
|
|||||
|
Distributed to:
* via CE Bulletin |
>Page 2 |
||||
| Contacts: Details of manufacturer contacts and MHRA contacts for technical and clinical aspects. Change of address or removal from address list for Healthcare Commission. |
>Page 3 |
||||
| Appendix: Smiths Medical customer letter, reply form, pre-use instructions and alternative methods of fluid warming (in pdf only). |
>Pages 4-10 |
|
Action deadlines for the Safety Alert Broadcast System (SABS)
|
|
|
Deadline (action underway): 30 August 2007
|
Deadline (action complete): 13 September 2007
|