Enter a keyword or phrase to search all pages in this section:


Advanced search

Archive:

Looking for a previously published document in this section?

MDA/2007/043 - Invacare Flamingo mobile patient hoists

Document details:

Type: Medical Device Alert
Series No: MDA/2007/043
Audience: Healthcare professionals
Published:
Format: Electronic and paper
Size: A4
Pages: 5
Price: Free
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Please note: In October and November 2010, the MHRA moved from its Market Towers address to a new location on Buckingham Palace Road. Please go to the contact us section of the website for updated contact details.

Page 1

Issued: 31 May 2007 at 11:00
Ref: MDA/2007/043

 

 
Immediate action
Tick mark
Action
 
Update
 
Information request


Device:
Invacare Flamingo mobile patient hoists.

 

Problem:
Detachment of spreader bar from hoist whilst in use.

>Page 2

Action by:
All those involved in the use and maintenance of mobile patient hoists, in particular nurses, care staff and maintenance staff or contractors.
 

Action:

  • Trace all Invacare Flamingo mobile patient hoists.
  • Carry out a visual check of the spreader bar attachment (see appendix).
  • If original design is fitted (see figure1) contact Invacare for a re-designed spreader bar kit, available free of charge.
  • Dispose of the old spreader bar following replacement.
 

Distributed to:

NHS trusts in England - Chief Executives*
Commission for Social Care Inspection (CSCI) - Headquarters
Healthcare Commission (CHAI) - Headquarters
Primary care trusts in England - Chief Executives*
Social services in England - Directors*

* via CE Bulletin

>Page 2

Contacts:
Details of manufacturer contacts and MHRA contacts for technical and clinical aspects.
Change of address or removal from address list for CSCI and Healthcare Commission.

>Page 3

Appendix:
Flamingo spreader bar attachment (original and new design).

>Page 5



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 02 July 2007
Deadline (action complete): 01 October 2007


Page last modified: 31 May 2007