| Issued: 23 May 2007 at 11:00 |
Ref: MDA/2007/040
|
|
|
Immediate action |
|
Action |
|
|
Update |
|
|
Information request |
| Device: Primus anaesthetic workstation supplied by Draeger Medical UK Ltd. |
>Page 2 |
||||||
| Problem: Potential misconnection of the backup oxygen cylinder; the oxygen cylinder inlet is situated alongside an oxygen outlet and both are fitted with the same connector. A misconnection to the oxygen outlet would prevent backup oxygen supply in the event of oxygen pipeline failure. |
>Page 2 |
||||||
|
Action by: |
|||||||
|
Action:
|
|||||||
|
Distributed to:
* via CE Bulletin |
>Page 2 |
||||||
| Contacts: Details of supplier contacts and MHRA contacts for technical and clinical aspects. Change of address or removal from address list for Healthcare Commission. |
>Pages 2-3 |
|
Action deadlines for the Safety Alert Broadcast System (SABS)
|
|
|
Deadline (action underway): 08 June 2007
|
Deadline (action complete): 22 June 2007
|