MDA/2007/038 - Bariatric profiling beds - Nightingale Pro Axis and Pro Axis Plus

Document details:

Type: Medical Device Alert
Series No: MDA/2007/038
Audience: Healthcare professionals
Published:
Format: Electronic and paper
Size: A4
Pages: 3
Price: Free
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Page 1

Issued: 21 May 2007 at 14:30
Ref: MDA/2007/038

 

 Tick mark
Immediate action
 
Action
 
Update
 
Information request


Device:
Bariatric profiling beds – Nightingale Pro Axis and Pro Axis Plus.

 

Problem:
Ignition of flammable gases from an overcharged battery within the main control box can result in the metal cover being blown off, putting those in the vicinity at risk of injury.
 

>Page 2  

Action by:
All those responsible for the maintenance of beds. This includes maintenance staff or contractors.

 

Action:

  • Trace all Nightingale Pro Axis and Pro Axis Plus beds manufactured before 01 April 2007.
  • Any of these beds known or suspected of having had their charging systems repaired or replaced should not be used until electrical checks have been carried out.
  • Contact the manufacturer on 01978 661 699 to arrange for these checks and subsequent modification of the beds.

 

>Page 2 

Distributed to:

NHS trusts in England - Chief Executives*
Commission for Social Care Inspection (CSCI) - Headquarters
Healthcare Commission (CHAI) - Headquarters
Primary care trusts in England - Chief Executives*
Social services in England - Directors


* via CE Bulletin

>Page 2

Contacts:
Details of manufacturer contacts and MHRA contacts for technical and clinical aspects.
Change of address or removal from address list for CSCI and Healthcare Commission.

>Page 3



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 31 May 2007
Deadline (action complete): 13 June 2007