MDA/2007/024 - All QUATTROcare spray cans, manufactured prior to 05 March 2007, for use in KaVo QUATTROcare dental handpiece maintenance systems

Document details:

Type: Medical Device Alert
Series No: MDA/2007/024
Audience: Healthcare professionals
Published:
Format: Electronic and paper
Size: A4
Pages: 4
Price:
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Please note: In October and November 2010, the MHRA moved from its Market Towers address to a new location on Buckingham Palace Road. Please go to the contact us section of the website for updated contact details.

Page 1

Issued: 15 March 2007 at 15:30
Ref: MDA/2007/024


Tick mark
Immediate action
 
Action
 Tick mark
Update
 
Information request


Device:
All QUATTROcare spray cans, manufactured prior to 05 March 2007, for use in KaVo QUATTROcare dental handpiece maintenance systems.

>Page 2

Problem:
KaVo Dental Ltd have recalled all their QUATTROcare spray cans due to a risk of fire/explosion. QUATTROcare spray cans may leak flammable gases when attached to the QUATTROcare maintenance system.

>Page 2

Action by:
All those involved in cleaning and maintaining dental handpieces
 

Action:

  1. Do not use sources of ignition (switches, naked flames, electrostatic charges etc.) in the environment where the QUATTROcare maintenance system is situated.
  2. Remove the QUATTROcare spray can from the maintenance system immediately.
  3. Check whether oil has leaked from the valve on the spray can. If oil is visible on the can’s valve ventilate the room immediately.
  4. Do not use the QUATTROcare maintenance system with spray cans that have been distributed prior to 05 March 2007. 
  5. Until an improved spray can is distributed by KaVo, clean and maintain dental handpieces using an alternative product as suggested by the manufacturer of the handpiece.
 

>Page 2

Distributed to:

NHS trusts in England - Chief Executives*
Healthcare Commission (CHAI) - Headquarters
Primary care trusts in England - Chief Executives*

* via CE Bulletin

>Page 2

Contacts:
Details of supplier contacts and MHRA contacts for technical and clinical aspects.
Change of address or removal from address list for Healthcare Commission.
 

 >Page 3

Appendix:
Manufacturer’s recall letter of 05 March 2007 (in pdf only).
 

 >Page 4



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 19 March 2007
Deadline (action complete): 26 March 2007


Page last modified: 15 March 2007