| Issued: 05 February 2007 at 14:00 |
Ref: MDA/2007/011
|
|
Immediate action |
|
|
Action |
|
Update |
|
|
Information request |
| This alert extends the guidance issued on 24 January 2007 in MDA/2007/007. | |||||
| Device: BOC Ltd, AZ size aluminium medical nitrous oxide cylinders fitted with a key operated pin index valve for use in MRI units. |
|||||
| Problem: Following the issue of MDA/2007/007, regarding AZ aluminium medical oxygen cylinders, BOC have informed MHRA that a similar problem has been found in a small number of medical nitrous oxide AZ cylinders. These cylinders have also been fitted with an obsolete pin index valve containing a ferromagnetic component that makes the cylinder potentially hazardous when used in an MRI environment. |
>Page 2 |
||||
| Action by: Anaesthetists, operating department practitioners and anaesthetic nurses working in MRI units. Hospital pharmacists, MRI safety advisors and MRI superintendent radiographers. |
|||||
|
Action:
|
Appendix | ||||
|
Distributed to:
* via CE Bulletin |
>Page 2 |
||||
| Contacts: Details of supplier contact and MHRA Defective Medicines contact. Change of address or removal from address list for Healthcare Commission. |
>Page 2 |
||||
| Appendix: Images of the affected valve and the two acceptable alternatives. |
>Pages 4-5 |
|