Page 1

Issued: 29 January 2007 at 14:00
Ref: MDA/2007/008


 
Immediate action
Tick mark
Action
 
Update
 
Information request


Device:
Neonatal radiant warmer: Draeger Resuscitaire, model RW82VHA-1C with variable height adjustment.

>Page 2

Problem:
Screws that secure the column to the base of the Resuscitaire may loosen over time, allowing the support column to tilt, due to an incorrect fitting.

>Page 2

Action by:
Neonatal special care units and EBME departments.
 

Action:

  • Identify Resuscitaire (model RW82VHA-1C) with an adjustable height mechanism.
  • If the unit is tilting, contact Draeger to arrange for your unit to be corrected.
  • Ensure that all in-house or third party service agents have access to Draeger’s Technical Service Bulletin (TSB) No. 01 that was issued on 21 December 2006.

>Page 2

Distributed to:

NHS trusts in England - Chief Executives*
Healthcare Commission (CHAI) - Headquarters

* via CE Bulletin

>Page 2

Contacts:
Details of manufacturer contacts and MHRA contacts for technical and clinical aspects.
Change of address or removal from address list for Healthcare Commission.

>Pages 2-3



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 26 February 2007
Deadline (action complete): 26 March 2007