| Issued: 17 January 2007 at 11:00 |
Ref: MDA/2007/005
|
|
Immediate action |
|
|
Action |
|
|
Update |
|
|
Information request |
| Device: Inditherm OTM1, OTM2, GTM1, RB1 & OTB1 operating theatre blankets and mattresses, used for maintaining patients’ body temperature. |
|||||
| Problem: Repeated folding of these devices can cause internal damage to the heating element. This can result in patient burns caused by localised hot-spots. The manufacturer is carrying out a modification which will detect this type of failure. This requires alteration of both the mattress and the control unit. |
|||||
| Action by: All those involved in the purchase, use or maintenance of these devices. |
|||||
|
Action:
|
|||||
|
Distributed to:
* via CE Bulletin |
>Page 2 |
||||
| Contacts: Details of manufacturer contacts and MHRA contacts for technical and clinical aspects. Change of address or removal from address list for Healthcare Commission. |
>Page 2 |
||||
| Appendix: Manufacturer’s advisory notice. |
>Page 4 |
|
Action deadlines for the Safety Alert Broadcast System (SABS)
|
|
|
Deadline (action underway): 24 January 2007
|
Deadline (action complete): 07 February 2007
|