MDA/2006/025 - Gambro Prisma haemodialysis sets: Prisma M60 Pre Set; Prisma M60 Set; Prisma M100 Pre Set; Prisma M100 Set; Prisma HF1000 Pre Set. Manufactured before the end of October 2005

Document details:

Type: Medical Device Alert
Series No: MDA/2006/025
Audience: Healthcare professionals
Published:
Format: Electronic and paper
Size: A4
Pages: 4
Price: free
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Page 1

Issued: 28 April 2006 at 11:00
Ref: MDA/2006/025

 

Tick mark
Immediate action
 
Action
 Tick mark
Update
 
Information request


Device:
Gambro Prisma haemodialysis sets:

  • Prisma M60 Pre Set
  • Prisma M60 Set
  • Prisma M100 Pre Set
  • Prisma M100 Set
  • Prisma HF1000 Pre Set

Manufactured before the end of October 2005.

 >Page 2

Problem:
During disconnection of the Prisma set from the Prisma haemodialysis system, blood can leak (under pressure) from the access pressure pod due to a manufacturing defect. Further to MDA/2005/066 which related to a limited recall of these devices, the manufacturer has initiated a recall of all batches produced before the end of October 2005.
 

 >Page 2

Action by:
Renal physicians, intensive care consultants and managers of supplies departments.
 

Action:

  • Immediately cease use of the affected lots, unless no alternative is available.
  • Locate and quarantine any products from affected lots if possible.
  • If only product from affected lots is available, follow the instructions for pre-use checks and disconnection described overleaf.
  • Ensure that the manufacturer is contacted to arrange collection and replacement of affected stock.
 

 >Pages 2-3

Distributed to:

NHS trusts in England - Chief Executives*
Healthcare Commission (CHAI) - Headquarters
Primary care trusts in England - Chief Executives*

* via CE Bulletin

 

 >Page 3

Contacts:
Details of manufacturer contacts and MHRA contacts for technical and clinical aspects.
Change of address or removal from address list for Healthcare Commission.
 

 >Page 3-4



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 12 May 2006
Deadline (action complete): 26 May 2006