Section search
Enter a keyword or phrase to search all pages in this section:
Advanced searchArchive:
Looking for a previously published document in this section?
| Type: | Medical Device Alert |
|---|---|
| Series No: | MDA/2012/004 |
| Audience: | Healthcare professionals |
| Published: | 9 February 2012 at 11.30 |
| Format: | Electronic |
| Size: | A4 |
| Pages: | 7 |
| Price: | Free |
| ISBN/ISSN: | |
| Author: | |
| Copyright: | Crown |
CAS deadlines
Action underway: 09 March 2012, action complete: 09 May 2012
Ref: MDA/2012/004 | Issued: 9 February 2012 at 11.30
If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2012/004 or 2011/011/018/401/009.
Technical aspects
Dr Rosalind Polley or Mrs Moji Ajeneye
Medicines & Healthcare products Regulatory Agency
Floor 4
151 Buckingham Palace Road
London SW1W 9SZ
Tel: 020 3080 7119 / 7271
Fax: 020 8754 3965
Email: rosalind.polley@mhra.gsi.gov.uk or mojisola.ajeneye@mhra.gsi.gov.uk
Clinical aspects
Nicola Lennard
Medicines & Healthcare products Regulatory Agency
Floor 4
151 Buckingham Palace Road
London SW1W 9SZ
Tel: 020 3080 7126
Fax: 020 8754 3965
Email: nicola.lennard@mhra.gsi.gov.uk
Please report via our website: Reporting adverse incidents involving medical devices
Further information about CAS can be found on the CAS website (external link)
Alerts in Northern Ireland will continue to be distributed via the NI SABS system.
Enquiries and adverse incident reports in Northern Ireland should be addressed to:
Northern Ireland Adverse Incident Centre
Health Estates Investment Group
Room 17
Annex 6
Castle Buildings
Stormont Estate
Dundonald
BT4 3SQ
Tel: 02890 523 704
Fax: 02890 523 900
Email: NIAIC@dhsspsni.gov.uk
Northern Ireland Adverse Incident Centre (NIAIC) (external link)
Please report directly to NIAIC, further information can be found on the NIAIC website (external link)
Further information about SABS can be found on the SABS website (external link)
Enquiries and adverse incident reports in Scotland should be addressed to:
Incident Reporting and Investigation Centre
Health Facilities Scotland
NHS National Services Scotland
Gyle Square
1 South Gyle Crescent
Edinburgh
EH12 9EB
Tel: 0131 275 7575
Fax: 0131 314 0722
Email: nss.iric@nhs.net
Health Facilities Scotland Incident Reporting and Investigation Centre (IRIC) (external link)
Reagents requiring manual handling for use in combination with cobas c501/502 analyser made by Hitachi.
Multiple tests are affected, such as syphilis (TPLA) test made by Sekisui.
All supplied by Roche.
Instrument
The cobas c501/ 502 laboratory analyser – manufactured by Hitachi and supplied by Roche – can run assays that are:
Cassette
In order to run partnership and some Roche assays on the cobas c501/502 instrument, users must manually transfer reagents from original bottles into a cobas c pack MULTI cassette (manufactured and supplied by Roche). They must also re-label the cassette with a new barcode supplied with the reagent. When the analyser reads this barcode the integrated cassette volume check (ICVC – the initial level check) function is not activated. This applies to all the assays listed below:
Assay
| Reagent | Assay | Manufacturer | Cat. No. | ICVC (initial level check) |
| CHE | Cholinesterase (Acetyl) | Roche Diagnostics | 11877763 216 | OFF |
| D-BIL | Direct Bilirubin | Roche Diagnostics | 11555413 216 | OFF |
| GLDH3 | Glutamate Dehydrogenase | Roche Diagnostics | 11929992 216 | OFF |
| HBDH | α-Hydroxybutyrate Dehydrogenase | Roche Diagnostics | 11876937 216 | OFF |
| TRIGB | Triglycerides | Roche Diagnostics | 11877771 216 | OFF |
| A1MG2 | Alpha-1 Microglobulin | Roche Diagnostics | 03576116 190 | OFF |
| B2MG | β 2-Microglobulin | Roche Diagnostics | 11660551 216 | OFF |
| KAPP2 | Kappa Light Chain | Roche Diagnostics | 11820176 316 | OFF |
| LAMB2 | Lambda Light Chain | Roche Diagnostics | 11820184 316 | OFF |
| MM (EDDP) | Methadone Metabolite | Thermo Fisher | 05178576 190 | OFF |
| OXY | Oxycodone | Thermo Fisher | 05179459 190 | OFF |
| TPLA | Syphilis | Sekisui | 04955137 190 | OFF |
Roche have updated the information encoded in the barcode of the syphilis (TPLA) assay such that the ICVC function will be activated for future lots, starting with lot no 628267.
Instructions for use
The instructions for use (IFU) for the affected assays on the c501/502 analyser need to be downloaded electronically via the TeleServiceNet to the customer cobas link. Instrument settings to run the syphilis and other assays on the cobas c501/502 instrument also have to be downloaded to the instrument via the cobas link. Note that the volumes described in the instrument settings should not be used for manual filling of reagent cassettes. Follow the reagent specific IFU for the cobas c501/ 502 instrument.
The use of insufficient volumes of reagent may lead to a false negative result leading to patient harm such as:
It is possible to obtain ‘valid’ but inaccurate results when there is no reagent remaining in the Roche cassette.
The MHRA has received a report of a known syphilis positive patient being reported as false negative using this combination of affected devices. Following a review of results the hospital identified the possibility of obtaining a ‘valid’ but inaccurate result where there is insufficient reagent, giving a false negative result.
The instrument settings downloaded via the cobas link contain instrument specific information for the application and do not contain customer instruction for reagent preparation. If these volumes are used reagent may deplete before the end of the run.
When running any of the assays listed above where level sensing is turned off it is possible to obtain an erroneous result without an error message.
MHRA advice on the use of IVDs in combination is published in DB2008(04).
Laboratory managers
Directors of pathology
GUM clinics
This MDA has been sent to:
This Device Alert is being sent to GPs for information only, in circumstances where patients may seek advice about the contents of this Notice. GPs need take no further action on receipt of this Alert.
Please bring this notice to the attention of relevant employees in your establishment. Below is a suggested list of recipients.
Trusts
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:
Primary care trusts
CAS liaison officers for onward distribution to all relevant staff including:
Health Protection Agency
Directors for onward distribution to:
Establishments registered with the Care Quality Commission (CQC) (England only)
This alert should be read by:
Please note: CQC and OFSTED do not distribute these alerts. Independent healthcare providers and social care providers can sign up to receive MDAs directly from the Department of Health’s Central Alerting System (CAS) by sending an email to: safetyalerts@dh.gsi.gov.uk and requesting this facility.