Ref. MDA/2005/043
Issued: 12 July 2005
| IMMEDIATE ACTION | |
| ACTION | ![]() |
| UPDATE | |
| INFORMATION REQUEST |
| Further Information | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| DEVICE: Styrene polymer based surgical and examination gloves - various manufacturers. |
> |
||||||||||
| PROBLEM: Some synthetic gloves degrade and may disintegrate when in contact with certain solvents, for example acetone, which is present in products such as spray-on wound dressings. |
> |
||||||||||
| ACTION BY: All users of latex-free surgical or examination gloves. |
|||||||||||
| ACTION: Ensure that styrene polymer based gloves do not come into contact with solvents. |
> |
||||||||||
|