Section search
Enter a keyword or phrase to search all pages in this section:
Advanced searchArchive:
Looking for a previously published document in this section?
| Type: | Medical Device Alert |
|---|---|
| Series No: | MDA/2010/064 |
| Audience: | Healthcare professionals |
| Published: | 12 August 2010 at 12:00 |
| Format: | Electronic only |
| Size: | A4 |
| Pages: | 5 |
| Price: | Free |
| ISBN/ISSN: | |
| Author: | |
| Copyright: | Crown |
Ref: MDA/2010/064 | Issued: 12 August 2010 at 12:00
If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2010/064 or 2010/007/023/081/016.
Technical aspects
Miss Feza Haque or Mr Ian Smith
Medicines and Healthcare products Regulatory Agency
Market Towers
1 Nine Elms Lane
London
SW8 5NQ
Tel: 020 7084 3066 / 3306
Fax: 020 7084 3106
Email: feza.haque@mhra.gsi.gov.uk and ian.smith@mhra.gsi.gov.uk
Clinical aspects
Dr Nicola Lennard
Medicines and Healthcare products Regulatory Agency
Market Towers
1 Nine Elms Lane
London
SW8 5NQ
Tel: 020 7084 3126
Fax: 020 7084 3111
Email: nicola.lennard@mhra.gsi.gov.uk
How to report adverse incidents
Please report via our website: Reporting adverse incidents involving medical devices.
Further information about CAS can be found on the Central Alerting System website (external link).
Alerts in Northern Ireland will continue to be distributed via the NI SABS system.
Enquiries and adverse incident reports in Northern Ireland should be addressed to:
Northern Ireland Adverse Incident Centre
Health Estates Investment Group
Room 17
Annex 6
Castle Buildings
Stormont Estate
Dundonald
BT4 3SQ
Tel: 02890 523 704
Fax: 02890 523 900
Email: NIAIC@dhsspsni.gov.uk
Northern Ireland Adverse Incident Centre (NIAIC) (external link)
How to report adverse incidents in Northern Ireland
Please report directly to NIAIC, further information can be found on the NIAIC website (external link).
Further information about SABS can be found on the SABS website (external link).
Enquiries and adverse incident reports in Scotland should be addressed to:
Incident Reporting and Investigation Centre
Health Facilities Scotland
NHS National Services Scotland
Gyle Square
1 South Gyle Crescent
Edinburgh
EH12 9EB
Tel: 0131 275 7575
Fax: 0131 314 0722
Email: nss.iric@nhs.net
Health Facilities Scotland - Incident Reporting and Investigation Centre (external link).
Enquiries in Wales should be addressed to:
Dr Sara Hayes
Senior Medical Officer
Medical Device Alerts
Welsh Assembly Government
Cathays Park
Cardiff
CF10 3NQ
Tel: 029 2082 3922
Email: Haz-Aic@wales.gsi.gov.uk
Novabel® dermal filler.
Manufactured by Merz Pharmaceuticals GmbH.
1x1 ml, Art.-Nr. 40800 and 2x1ml, Art.-Nr. 49021.
All batches.
Novabel® is a colourless dermal filler composed of cross-linked alginate. It is supplied in a prefilled, 1 ml, single-use syringe with two sterile 30G ½ needles.
Use of Novabel® dermal filler may cause adverse reactions.
Merz Pharmaceuticals GmbH issued a Field Safety Notice (dated 23 July 2010), advising practitioners to stop using Novabel®.
The manufacturer has received reports of adverse reactions to the filler including redness, bruising, pain, swelling and histologically confirmed granuloma. They have also received reports of nodules and indurations in the infra-orbital area.
This MDA has been distributed to:
Onward distribution
Please bring this notice to the attention of all who need to know or be aware of it. This may include distribution by:
Trusts to:
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:
Care Quality Commission (CQC) (England only) to:
The MHRA considers this information to be important to:
Health Protection Agency to:
Directors for onward distribution to:
Primary care trusts to:
CAS liaison officers for onward distribution to all relevant staff including:
Medical Information Department
Merz Pharma UK Ltd
260 Centennial Park
Elstree Hill South
Elstree
Herts
WD6 3SR
Tel: 020 8236 0000
Fax: 020 8236 3501
Email: medical.information@merz.com