Ref: MDA/2010/029 | Issued: 13 April 2010 at 15:00
If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2010/029 or 2009/011/003/401/001.
Jonathan Smith or Emma Rooke
Medicines & Healthcare products Regulatory Agency
Market Towers
1 Nine Elms Lane
London
SW8 5NQ
Tel: 020 7084 3600
Fax: 020 7084 6177
Email: bav@mhra.gsi.gov.uk
How to report adverse incidents
Please report via our website: Reporting adverse incidents involving medical devices.
Further information about CAS can be found on the Central Alerting System website (external link).
Enquiries and adverse incident reports in Northern Ireland should be addressed to:
Northern Ireland Adverse Incident Centre
Health Estates Investment Group
Room 17
Annex 6
Castle Buildings
Stormont Estate
Dundonald
BT4 3SQ
Tel: 02890 523 704
Fax: 02890 523 900
Email: NIAIC@dhsspsni.gov.uk
Northern Ireland Adverse Incident Centre (NIAIC) (external link)
How to report adverse incidents in Northern Ireland
Please report directly to NIAIC, further information can be found on the NIAIC website (external link).
Further information about SABS can be found on the SABS website (external link).
Enquiries and adverse incident reports in Scotland should be addressed to:
Incident Reporting and Investigation Centre
Health Facilities Scotland
NHS National Services Scotland
Gyle Square
1 South Gyle Crescent
Edinburgh
EH12 9EB
Tel: 0131 275 7575
Fax: 0131 314 0722
Email: nss.iric@nhs.net
Health Facilities Scotland - Incident Reporting and Investigation Centre (external link).
Enquiries in Wales should be addressed to:
Dr Sara Hayes
Senior Medical Officer
Medical Device Alerts
Welsh Assembly Government
Cathays Park
Cardiff
CF10 3NQ
Tel: 029 2082 3922
Email: Haz-Aic@wales.gsi.gov.uk

Mobile hoists.
Arjo Minstrel standard type HMA001 and EPC type HMA0011 manufactured by Medibo Medical NV.
Models supplied after December 2005 with a serial number as shown in the list in the appendix. The serial number is located on the hoist identification label on the lower section of the mast. Arjo Minstrel hoists fitted with Arjo weigh scales are not affected as they use a different attachment method.
Inaccurate maintenance instructions and poor inspection of hanger bar attachment. Attachments have failed in use leading to a fall and injury to the person being moved.
There have been several incidents where the spreader bar has detached from the hoist boom due to failure of the attachment pivot bolt.
The bolt is welded to the hanger bar and cannot be serviced or replaced. The preventative maintenance schedule for the hanger bar, as detailed in the Minstrel operating and product care instructions (issue 4), is incorrect in requiring replacement of the bolt every three years by Arjo personnel. It should state that the complete spreader bar needs to be replaced every three years. The instructions also lack detail on the frequent checks required for the spreader bar attachment during the three year period.
Medibo Medical NV is updating its instructions and redesigning the hanger bar attachment but the new design is not yet available. Existing hoists require inspection to check if they are safe to use until the new design is available. Arjo has compiled a Field Safety Notice (FSN) giving detail of the inspection required by a competent person. However, this has not yet reached all users.
This MDA has been distributed to:
Onward distribution
Please bring this notice to the attention of all who need to know or be aware of it. This may include distribution by:
Trusts to:
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:
Care Quality Commission (CQC) (England only) to:
The MHRA considers this information to be important to:
OFSTED to:
The MHRA considers this to be important to:
Primary care trusts to:
CAS liaison officers for onward distribution to all relevant staff including:
Social services to:
Liaison officers for onward distribution to all relevant staff including:
Arjo Med AB Ltd
Customer Services
Tel: 07807 054713
Fax: 01452 428344
Email: uksales@arjo.co.uk
If you have any comments or feedback on this Medical Device Alert, please email us on: dts@mhra.gsi.gov.uk