| Type: | Medical Device Alert |
|---|---|
| Series No: | MDA/2010/010 |
| Audience: | Healthcare professionals |
| Published: | 1 February 2010 at 11:00 |
| Format: | Electronic only |
| Size: | A4 |
| Pages: | 4 |
| Price: | Free |
| ISBN/ISSN: | |
| Author: | |
| Copyright: | Crown |
Please note: In October and November 2010, the MHRA moved from its Market Towers address to a new location on Buckingham Palace Road. Please go to the contact us section of the website for updated contact details.
Ref: MDA/2010/010 | Issued: 1 February 2010 at 11:00
If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2010/010 or 2009/006/003/291/015.
James McVeigh or Ed Ball
Medicines & Healthcare products Regulatory Agency
Centre for Assistive Technology
241 Bristol Avenue
Bispham
Blackpool
FY2 0BR
Tel: 01253 596 000
Fax: 01253 596 177
Email: bav@mhra.gsi.gov.uk
How to report adverse incidents
Please report via our website: Reporting adverse incidents involving medical devices.
Further information about CAS can be found on the Central Alerting System website (external link).
Alerts in Northern Ireland will continue to be distributed via the NI SABS system.
Enquiries and adverse incident reports in Northern Ireland should be addressed to:
Northern Ireland Adverse Incident Centre (NIAIC)
Health Estates
Estate Policy Directorate
Stoney Road
Dundonald
Belfast
BT16 1US
Tel: 02890 523 704
Fax: 02890 523 900
Email: NIAIC@dhsspsni.gov.uk
Northern Ireland Adverse Incident Centre (NIAIC) (external link)
How to report adverse incidents in Northern Ireland
Please report directly to NIAIC, further information can be found on the NIAIC website (external link).
Further information about SABS can be found on the SABS website (external link).
Enquiries and adverse incident reports in Scotland should be addressed to:
Incident Reporting and Investigation Centre
Health Facilities Scotland
NHS National Services Scotland
Gyle Square
1 South Gyle Crescent
Edinburgh
EH12 9EB
Tel: 0131 275 7575
Fax: 0131 314 0722
Email: nss.iric@nhs.net
Health Facilities Scotland - Incident Reporting and Investigation Centre (external link)
Enquiries in Wales should be addressed to:
Dr Sara Hayes
Senior Medical Officer
Medical Device Alerts
Welsh Assembly Government
Cathays Park
Cardiff
CF10 3NQ
Tel: 029 2082 3505 / 3922
Email: Haz-Aic@wales.gsi.gov.uk
C-Max U/2 stair climbers manufactured by Alber Antriebstechnik GmbH (AAT).
This problem affects C-Max U/2 stair climbers which have:
The serial number is displayed on a silver sticker on the motor unit of the stair climber.
Other models of AAT stair climber are not affected by this notification as they have a different drive system design.
There is a risk of injury to the attendant’s hand, by the climbing mechanism, if the stair climber is activated when two people are moving it.
The MHRA was made aware of an incident in which an attendant was injured whilst the (unoccupied) stair climber was being handled by two people. One attendant’s thumb was trapped against the lower frame tube by the climbing mechanism.
As a result of this incident the manufacturer has:
AAT issued a Field Safety Notice in October 2009 but have been unable to trace all existing stair climbers.
1. Trace all affected stair climbers.
2. Contact AAT to arrange for a protective cover to be fitted around the mechanism and to obtain a copy of the revised instructions for use.
3. Ensure that all operators and attendants are aware of these revised instructions for use.
This MDA has been distributed to:
Onward distribution
Please bring this notice to the attention of all who need to know or be aware of it. This may include distribution by:
Trusts to:
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:
Care Quality Commission (CQC) (England only) to:
The MHRA considers this information to be important to:
OFSTED to:
The MHRA considers this to be important to:
Primary care trusts to:
CAS liaison officers for onward distribution to all relevant staff including:
Social services to:
Liaison officers for onward distribution to all relevant staff including:
Supplier:
Mr C Singh
AAT (GB) Ltd
Unit 6, Miners Road
Llay Industrial Estate
Llay
Wrexham
LL12 0JP
Tel: 01978 853 900
Fax: 01978 853 930
Email: c.singh@aatgb.com
If you have any comments or feedback on this Medical Device Alert, please email us at: dts@mhra.gsi.gov.uk