Section search
Enter a keyword or phrase to search all pages in this section:
Advanced searchArchive:
Looking for a previously published document in this section?
| Type: | Medical Device Alert |
|---|---|
| Series No: | MDA/2010/007 |
| Audience: | Healthcare professionals |
| Published: | 18 January 2010 at 15:00 |
| Format: | Electronic only |
| Size: | A4 |
| Pages: | 5 |
| Price: | Free |
| ISBN/ISSN: | |
| Author: | |
| Copyright: | Crown |
Please note: In October and November 2010, the MHRA moved from its Market Towers address to a new location on Buckingham Palace Road. Please go to the contact us section of the website for updated contact details.
Ref: MDA/2010/007 | Issued: 18 January 2010 at 15:00
If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2010/007 or 2009/009/018/401/015.
Technical aspects
Nicole Small or Richard Glover
Medicines and Healthcare products Regulatory Agency
Market Towers
1 Nine Elms Lane
London
SW8 5NQ
Tel: 020 7084 3310 / 3245
Fax: 020 7084 3209
Email: nicole.small@mhra.gsi.gov.uk or richard.glover@mhra.gsi.gov.uk
Clinical aspects
Jonathan Plumb
Medicines and Healthcare products Regulatory Agency
Market Towers
1 Nine Elms Lane
London
SW8 5NQ
Tel: 020 7084 3128
Fax: 020 7084 3111
Email: jonathan.plumb@mhra.gsi.gov.uk
How to report adverse incidents
Please report via our website: Reporting adverse incidents involving medical devices.
Further information about CAS can be found on the Central Alerting System website (external link).
Alerts in Northern Ireland will continue to be distributed via the NI SABS system.
Enquiries and adverse incident reports in Northern Ireland should be addressed to:
Northern Ireland Adverse Incident Centre (NIAIC)
Health Estates
Estate Policy Directorate
Stoney Road
Dundonald
Belfast
BT16 1US
Tel: 02890 523 704
Fax: 02890 523 900
Email: NIAIC@dhsspsni.gov.uk
Northern Ireland Adverse Incident Centre (NIAIC) (external link)
How to report adverse incidents in Northern Ireland
Please report directly to NIAIC, further information can be found on the NIAIC website (external link).
Further information about SABS can be found on the SABS website (external link).
Enquiries and adverse incident reports in Scotland should be addressed to:
Incident Reporting and Investigation Centre
Health Facilities Scotland
NHS National Services Scotland
Gyle Square
1 South Gyle Crescent
Edinburgh
EH12 9EB
Tel: 0131 275 7575
Fax: 0131 314 0722
Email: nss.iric@nhs.net
Health Facilities Scotland - Incident Reporting and Investigation Centre (external link)
Enquiries in Wales should be addressed to:
Dr Sara Hayes
Senior Medical Officer
Medical Device Alerts
Welsh Assembly Government
Cathays Park
Cardiff
CF10 3NQ
Tel: 029 2082 3922
Email: Haz-Aic@wales.gsi.gov.uk
All pole clamps for Infusomat Space and Perfusor Space infusion pumps.
Manufactured by B Braun.
Part number 8713130.
This MDA has been distributed to:
Onward distribution
Please bring this notice to the attention of all who need to know or be aware of it. This may include distribution by:
Trusts to:
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:
Care Quality Commission (CQC) (England only) to:
The MHRA considers this information to be important to:
Primary care trusts to:
CAS liaison officers for onward distribution to all relevant staff including:
Quality Team
B Braun Medical Ltd
Thorncliffe Park
Sheffield
S35 2PW
Tel: 0114 225 9155
Fax: 0114 225 9136
Email: catherine.clulow@bbraun.com