| Issued: 24 February 2009 at 12:00 |
Ref: MDA/2009/015
|
|
|
Immediate action |
|
|
Immediate action update |
|
Action |
|
|
Action update |
| Device: All Action 2000 manual wheelchairs manufactured by Invacare with serial numbers from 7007533957 to 7008568013 inclusive. |
| Problem: The bolts used to secure the castor assemblies to the wheelchair frame can fail in use. This has caused occupants to fall from their wheelchair and sustain injuries. |
|
Action:
|
| Action by: All those involved in the provision, prescription, repair, maintenance and use of these wheelchairs. In particular: wheelchair service managers, rehabilitation engineers, occupational therapists, maintenance staff and contractors. |
|
Contact: Manufacturer E-mail: uk@invacare.com |
|
Action deadlines for the Central Alerting System (CAS)
|
|
|
Deadline (action underway): 24 March 2009
|
Deadline (action complete): 27 May 2009
|