MDA/2008/080 - Enteral feeding pump - ClearStar model M771 manufactured by Abbott Nutrition

Document details:

Type: Medical Device Alert
Series No: MDA/2008/080
Audience: Healthcare professionals
Published: 12 November 2008
Format: Electronic and paper
Size: A4
Pages: 3
Price: Free
ISBN/ISSN: n/a
Author:
Copyright: Crown
   

Help viewing PDFs:

Page 1

Issued: 12 November 2008 at 11:00
Ref: MDA/2008/080

 

 
Immediate action
Tick mark
Action
 
Update
 
Information request

 

Device:
Enteral feeding pump. ClearStar model M771 manufactured by Abbott Nutrition. Specific serial numbers.

 

Problem:
Risk of under-infusion due to cracking/breaking of an internal adapter bracket.

Following an initial service programme of these pumps in 2007 (see MDA/2007/060), Abbott has identified another 298 pumps that may be affected. Abbott is running a second service programme but is unable to locate 38 of these pumps (serial numbers listed in the appendixPDF file (opens in new window) (629Kb)).

 

Action by:
All staff responsible for prescribing and maintaining enteral feeding devices, particularly dietetic departments and EBME departments.

 

Action:

  • Check the list of serial numbers in the appendix. Contact Abbott to arrange exchange if the pump is listed.
  • Ensure that users are aware of the manufacturer's customer notice. Where the location of the pump is known to Abbott, its customers will be contacted by Abbott (see appendix).

 

Distributed to:

 

NHS trusts in England

– Chief Executives*  

OFSTED – Directors of Children’s Services
Commission for Social Care Inspection (CSCI) – Headquarters
Healthcare Commission (CHAI) – Headquarters
Primary care trusts in England – Chief Executives*
NHS Boards in Scotland – Chief Executives

* via CE Bulletin 

►Page 2

Contacts:
Details of manufacturer contacts and contacts for technical and clinical aspects.

Change of address or removal from address list for CSCI and Healthcare Commission.

►Page 3

Appendix:
Abbott Nutrition customer notice and list of serial number of pumps where location is unknown.

PDF OnlyPDF file (opens in new window) (629Kb)

 

Action deadlines for the Central Alerting System (CAS)
Deadline (action underway): 10 December 2008
Deadline (action complete): 21 January 2009