Device:
Gambro and Hospal blood sets (lot numbers 0806 – 0817).

> Page 2

Problem:
Gambro has identified the potential for a total or partial occlusion of the arterial (red) and/or venous (blue) dialyser connectors of these blood tubing sets. This is due to a moulding defect.
 
Action by:
All renal unit staff.
 

Action:
Verify if product (box/single package) has a green dot on packaging.
If no green dot is present on the blood tubing sets:

  • remove set from packaging and visually inspect sets prior to use to check for occlusions
  • if inspection shows set is occluded DO NOT USE THE SET
  • quarantine device and contact Gambro for replacement.
 

> Page 2

Distributed to:

NHS trusts in England - Chief Executives*
Healthcare Commission (CHAI) - Headquarters
NHS Boards in Scotland - Chief Executives

* via CE Bulletin

> Page 2

Contacts:
Details of manufacturer contacts and contacts for technical and clinical aspects. Change of address or removal from address list for Healthcare Commission.

> Page 3

Appendix:
Gambro Customer Communication: Follow-up notice.

 PDF Version Only



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 03 July 2008
Deadline (action complete): 21 July 2008