| Issued: 16 June 2008 at 11:30 |
Ref: MDA/2008/039
|
|
Immediate action |
|
|
Action |
|
|
Update |
|
|
Information request |
| Device: Medtronic Physio-Control LIFEPAK® 12, LIFEPAK® 20 and LIFEPAK® 20e external defibrillators with certain settings. |
>Page 2 |
||||||||||
| Problem: When used in AED mode with ‘Auto Analyze’ ON, the ECG analysis (Shock Advisory System™) may start before warning is given to the user to stand clear of the patient. This can result in inappropriate shock or no shock recommendations being made by the device due to CPR or movement artefacts. |
|
||||||||||
| Action by: All those involved in the use or maintenance of these devices. |
|
||||||||||
|
Action:
|
>Page 2 |
||||||||||
|
Distributed to:
* via CE Bulletin |
>Page 2 |
||||||||||
| Contacts: Details of manufacturer contacts and contacts for technical and clinical aspects. Change of address or removal from address list for CSCI and Healthcare Commission. |
>Page 3 |
||||||||||
| Appendix: Instructions for changing ‘Auto Analyze’ setup options (in pdf only). |
>Pages 5-6 |
|
Action deadlines for the Safety Alert Broadcast System (SABS)
|
|
|
Deadline (action underway): 23 June 2008
|
Deadline (action complete): 07 July 2008
|