MDA/2008/021 - Hospital beds Eleganza Standard, Smart and Deluxe models supplied by Pegasus

Document details:

Type: Medical Device Alert
Series No: MDA/2008/021
Audience: Healthcare professionals
Published:
Format: Electronic and paper
Size: A4
Pages: 7
Price: Free
ISBN/ISSN:
Author:
Copyright: Crown
   

Help viewing PDFs:

Please note: In October and November 2010, the MHRA moved from its Market Towers address to a new location on Buckingham Palace Road. Please go to the contact us section of the website for updated contact details.

Page 1

Issued: 27 March 2008 at 12:00
Ref: MDA/2008/021
 
Immediate action
Tick mark
Action
 
Update
 
Information request


Device:
Hospital beds Eleganza Standard, Smart and Deluxe models supplied by Pegasus.

› Page 2

Problem:
The emergency backrest release function on these beds may fail or become stiff and difficult to operate leading to a risk of delay in performing CPR.

› Page 2

Action by:
All those using, providing and maintaining hospital beds. In particular bed managers, nurses, care staff, maintenance staff and contractors.
 

Action:

  • Trace all Eleganza Standard, Smart and Deluxe beds that are in use or in storage.
  • From these identify beds that are fitted with Dewert actuators (see page 2).
  • Check that the electrical and manual CPR functions are operating correctly (see page 2).
  • Contact Linet UK to arrange for all Dewert backrest actuators to be exchanged (see contact on page 3).
 

› Page 2

Distributed to:

NHS trusts in England - Chief Executives*
Commission for Social Care Inspection (CSCI) - Headquarters
Healthcare Commission (CHAI) - Headquarters
Primary care trusts in England - Chief Executives*
Social services in England - Directors*

* via CE Bulletin

› Page 3

Contacts:
Details of manufacturer contacts and MHRA contacts for technical and clinical aspects. Change of address or removal from address list for CSCI and Healthcare Commission.

› Pages 3-4

Appendix (pdf only):
List of model and serial numbers for affected beds (in pdf only).

› Pages 5-7



Action deadlines for the Safety Alert Broadcast System (SABS)
Deadline (action underway): 28 April 2008
Deadline (action complete): 27 June 2008


Page last modified: 27 March 2008